Tempmate M1 – Multi Use Logger
Cost-effective multi-temperature data logger with automatic PDF-evaluation
Features
USB PortIP 67 protected
Measuring range: -30 ° C to + 70°C
Output as PDF and CSV Report
GDP conformity
-
Monitoring of the cold chain - It's easy with temp mate . ® -M1 temperature data loggers.
temp mate . ® -M1 datalogger have been developed specifically for monitoring temperature-sensitive products throughout the cold chain. Thanks to the custom factory settings are temp mate . ® -M1 datalogger immediately ready to record all relevant data cooling.Automatic PDF report
The automatically generated PDF report contains detailed information on the assessment of the monitored goods: data curve, statistical values such as MIN, MAX, AVG, MKT and each measurement value table.The benefits of our temperature data logger at a glance
Traceable recording of temperature variations within a cold chain
Cost-effective and reliable
Very light and small: can attach virtually any packaging
Reusable
No additional software needed to evaluate
Detailed analysis with chart and table
Display shows status
The innovative reusable data logger temp mate . ® -M1 with integrated USB port is the optimum data logger for cost temperature monitoring of stationary refrigeration equipment or transportation.
For evaluation, the temp mate . ® -M1directly connected to the USB interface. Here a PDF report is automatically generated, which provides in addition to the temperature profile all individual values and statistical information. Hardware (cable interface) or software are not required.
To adjust the data logger or evaluation of the measured data is the free temp base . ® -D V2 software required. -
datalogger type Disposable data logger temperature range -30 ° C to + 70 ° C accuracy ± 0.5 ° C (at -20 ° C to + 40 ° C) resolution 0.1 ° C memory size 32,000 measurements battery CR2032 / replaceable by customer measuring interval 10 Sec. up to 24 hours. recording time By customer programmable / up to 12 months start method Manually by pressing a button or automatically at the preset time alarm type Single alarm or cumulative protection class IP67 dimension 79 mm x 33 mm x 14 mm (LxWxH) weight 25g certified EN12830, validation certificate, CE, RoHS software PDF Reader autom. PDF report Yes PC connection USB 2.0, Type A -
Are temp mate . ® -M1 Logger IATA compliant and can thus be used on airplanes?
Yes, temp mate . ® -M1 reusable data logger working with a 3V lithium battery as a power source, which may be used in accordance with IATA in airplanes. When shipping, please bring the following documents to your shipment: Certification for Safe Transport of Chemical-Goods Lithium Battery-LabelHow accurate is the built-up real-time clock (RTC)?
The real time clock of the logger proceeds of production. For technical reasons, there may be a deviation of ± 2 minutes after 5 months or ± 4 minutes after 10 months.How long does the battery last?
Each battery will run continuously for about half a year, the lifetime results from the measurement interval and from the storage and use. The battery is customer replaceable.If the temp mate . ® -M1 calibrated logger?
Yes, with each delivery, you will receive a free certificate.Can temp mate . ® evaluate -M1 logger with software and export the data?
The temp mate . ® -M1 is optional with the temp base . ® -D V2 compatible software. To ensure that all data can be managed and z. B. be exported as PDF or CSV file. The temp base . ® -D V2 software is available here or in the right box under software download free download.What is the evaluation of the PDF report?
The temperature is shown as a graph, as well as all individual values in tables. In addition, further statistical information (eg serial number) are included.Why can not I open the PDF report?
You need Adobe Acrobat version 9.0 or higher to open the PDF. The latest version can be downloaded here:If the PDF report is protected against manipulation?
Like any electronic document, a PDF file can be manipulated. However, even after the data evaluation, the original data remains on the temp mate . ® -M1 logger stored so that they can be downloaded again and again again and compared with the questionable PDF report. Therefore it is important that until the final clarification of the logger are repealed.Will also between the measurement intervals, the temperature monitored?
The temperature is also measured between the measuring intervals and each formed an average value thereof. A significant temperature variation of ± 2 ° C, this is recorded accordingly.What time is specified in the evaluation?
The temp mate . ® -M1 dataloggers have integrated a real time clock (RTC). This is factory set to GMT ± 0 and is therefore used for PDF-evaluating the temp mate . ® used -M1. If the temp mate . ® However -M1 data logger with the optional temp base . ® -D V2 read, the software converts the reports according to the local computer system time.