Tempmate S1 – Single Use Logger
Disposable temperature logger with automatic PDF and CSV report and 2 years shelf life
Features
USB PortIP 67 protected
Measuring range: -30 ° C to + 70°C
Output as PDF and CSV Report
GDP conformity
-
Monitoring of the cold chain - It's easy with temp mate . ® -S1 V2
temperature data loggers.
temp mate . ® -S1 V2 data loggers were developed specifically for monitoring temperature-sensitive products throughout the cold chain. Thanks to the custom factory settings are temp mate . ® -S1 V2 datalogger immediately ready to record all relevant data cooling.Automatic PDF and CSV report
The PDF automatically generated and CSV reports contain detailed information about the evaluation of the monitored goods: data curve, statistical values such as MIN, MAX, AVG, MKT and each measurement value table.The benefits of our temperature data logger at a glance
2 years shelf life
Automatic PDF and CSV report
Up to 180 days recording time
reprogrammable
Up to 5 alarm limits
Backup log protection mechanism
Traceable recording of temperature variations within a cold chain
Cost-effective and reliable
Very light and small: can attach virtually any packaging
Cover: waterproof, food safe
No additional software needed
Detailed analysis with chart and table
LED lights indicate current status
The innovative disposable data logger temp mate . ® -S1 V2 with integrated USB port is the ideal companion for goods cost temperature monitoring of transports.
The temp mate . ® -S1 V2 logger is programmed and shipped in a prior configuration. If necessary, the logger by customer can be reprogrammed and re-sealed. The following settings can temp base . ® -D V2 software are made:
Maturity to 180 days
Up to 5 alarm thresholds
password protection
time zone
Alarm time delay (default: 30 min.)
Start time delay (default: 0 min.)
The loggers are immediately usable and can be programmed by the customer if required. For evaluation, the temp mate . ® -S1 V2 connected directly to the USB port. Here a PDF and CSV report is automatically generated, which provides in addition to the temperature profile all individual values and statistical information. Hardware (cable interface) or software are not required. -
datalogger type Disposable data logger temperature range -30 ° C to + 70 ° C (-22 ° F to 158 ° F) accuracy ± 0.5 ° C / 0.9 ° F (-20 ° C / + 40 ° C) ± 1.0 edge regions resolution 0.1 ° C / ° F memory size 16,000 measurements durability / battery Including 2 years / CR2450 button cell battery measuring interval 10 minutes (standard - others on request) recording time 110 days (standard - others on request) start method Manual start or temp base . ® -D V2 stop method Manual stop, temp base . ® -D V2 or automatically when full alarm type Manual stop or automatically after 90 days protection class IPE67 / NEMA 6 dimension 88mm x 47mm x 7mm [LxWxH] weight about 14.6g certified EN12830, CE, RoHS validation certificate Expression and in the memory of the logger software PDF or CSV Reader / Optional temp base . ® -D V2 automatic report PDF and CSV PC connection USB 2.0, Type A alarm configuration Optional up to 5 limits programmable Optional on temp base . ® -D V2 delivery Shipping documents, certificate, inscription label and seal bags -
Are temp mate . ® -S1 V2 Logger IATA compliant and can thus be used on airplanes?
Yes, temp mate . ® -S1 V2 Disposable data logger working with a 3V lithium battery as a power source, which may be used in accordance with IATA in airplanes. When shipping, please bring the following documents to your shipment: Certification for Safe Transport of Chemical-Goods and Lithium Battery-LabelHow accurate is the built-up real-time clock (RTC)?
The real time clock of the logger proceeds of production. For technical reasons, there may be a deviation of ± 2 minutes after 5 months or ± 4 minutes after 10 months.As long as the loggers are Stable / storable?
Each logger is factory least 1 year shelf life, a maximum of 24 months. The respective MHD is printed on the back and should be considered STEST.If the temp mate . ® -S1 V2 calibrated logger?
Yes, with each delivery, you will receive a free certificate.Can temp mate . ® -S1 V2 evaluate logger with software and export the data?
The temp mate . ® -S1 V2 is optional with the temp base . ® -D V2 compatible software. To ensure that all data can be managed and z. B. be exported as PDF or CSV file. The temp base . ® -D V2 software is available here or in the right box under software download free download.What is the evaluation of the PDF report?
The temperature is shown as a graph, as well as all individual values in tables. In addition, further statistical information (eg serial number) are included.Why can not I open the PDF report?
You need Adobe Acrobat version 9.0 or higher to open the PDF. The latest version can be downloaded here:If the PDF report is protected against manipulation?
Like any electronic document, a PDF file can be manipulated. However, even after the data evaluation, the original data remains on the Logger stored so that they can be repeatedly re-downloaded and synchronized with the questionable PDF report. Therefore it is important that until the final clarification of the logger are repealed.Will also between the measurement intervals, the temperature monitored?
The temperature is also measured between the measuring intervals and each formed an average value thereof. A significant temperature variation of ± 2 ° C, this is recorded accordingly.What time is specified in the evaluation?
The temp mate . ® -S1 V2 dataloggers have integrated a real time clock (RTC). This is factory set to GMT ± 0 and is therefore used for PDF-evaluating the temp mate . ® -S1 V2 used. The start time is in this case the exact second documented and stop time always corresponds to the last measurement interval during which it was stopped.